Common Faults of Salt Spray Environmental Chambers
The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.
Salt spray environmental chamber for sale
The salt spray environmental chamber cannot be heated
1. Heat the solid state relay, if there is no output, replace the solid state relay.
2. If the external dial has no output, replace the external dial.
3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.
4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.
5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.
The heating temperature is abnormal
1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.
2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.
3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.
4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.
The external dial doesn’t display
1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.
2. Check whether or not the power switch is damaged. If damaged, replace the power switch.
3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.
4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.
5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.
If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.
Recomendações relacionadas
-
What should I pay attention to when replacing a water-cooled chiller compressor
998Once the compressor of the water-cooledchiller fails, the compressor needs to be replaced in time, but the replacementof the compressor is not as simple as imagined.Before the water-cooled chiller compressoris replaced, the reasons for the dama...
Ver pormenores -
Precauções para a utilização do refrigerador de recirculação da LNEYA
1130LNEYA's low temperature recirculating chillercan be used with supporting instruments (such as rotary evaporators,fermentation tanks, chemical reactors, freeze-drying equipment,biopharmaceutical reactors, etc.), with large cooling capacity and f...
Ver pormenores -
Que aspectos devem ser considerados no sistema de arrefecimento por circulação externa do banco de ensaio?
1019Como todos sabemos, a sala de ensaios tem de estar equipada com um sistema de arrefecimento para o desempenho do veículo e transmissão abrangente, simulação de estrada, compatibilidade electromagnética, bateria de energia, ambiente do veículo e outros ensaios. No sector automóvel...
Ver pormenores -
Maintenance of the condenser of high and low temperature circulation device
1144The condenser in the high and low temperature constant temperature circulation device is one of the accessories in the refrigeration process, and its performance is very important. LNEYA reminds the user that regular maintenance is required after ...
Ver pormenores