The structure of the scroll compressor and the piston compressor in the low-to-86 degree low-temperature refrigerator is different, and it cannot be reversed. The reverse phase of the three-phase power supply of the low-zero-86 degree low-temperature refrigerator will cause the compressor to reverse, so it is necessary to install reverse phase protection. To prevent the refrigeration press from reversing. After the reverse phase protector is installed, the compressor can be operated in the normal phase and does not operate in the reverse phase. When the sub-zero-86 degree low temperature refrigerator is reversed, just reverse the two lines of the power supply and change to the normal phase.
In order to protect the compressor under high load operation or insufficient refrigerant, an exhaust temperature protector needs to be installed in the subzero-86 degree low temperature refrigerator system, and the exhaust gas temperature is set within the range of 130±5°C. The compressor is stopped. This temperature value refers to the position detection temperature of the compressor exhaust pipe within 10 cm from the outlet.
In order to protect the operation of the subzero-86 degree low temperature refrigerator compressor when the refrigerant is insufficient, a low pressure switch is required. When the temperature is set to 0.03 mpa or more, the compressor stops running. Zero-to-86 degree low-temperature refrigerator compressors operate in the state of insufficient refrigerant, then the temperature of the compressor part and the motor part of the sub–86 degree low-temperature refrigerator will rise instantaneously. At this time, the low-voltage switch of the lower-86 degree low-temperature refrigerator can be Protection against damage to the compressor and motor burnout that cannot be protected by the internal temperature and exhaust temperature protectors.
In order to prevent the zero-to-86 degree low-temperature refrigerator from running 24 hours, the high-load operation of the lower-86 degree low temperature refrigerator compressor, the electromagnetic switch is poor, the overcurrent caused by the holding shaft, or the motor is burnt due to the motor temperature. The internal thermostat is installed inside the compressor. It is installed on the neutral contact of the three-phase motor. When the subzero-86 degree low temperature refrigerator is abnormal, the motor is protected by cutting off the three phases at the same time.
The zero-to-86 degree low-temperature refrigerator electromagnetic switch is used to control the compressor operation of the zero-86 degree low-temperature refrigerator. The shutter is closed for the purpose of installation. When installing, it should be kept vertical. If the installation is wrong, the joint spring pressure will change and noise will be generated. It causes phase loss operation. For models with a compressor with a direct power failure protector, there is no need to load the protector.
The protection device of the minus-86 degree low temperature refrigerator is very necessary. All the users are advised to choose the time, if it is capable, Wuxi Guanya LNEYA recommends the explosion-proof zero-86 degree low temperature refrigerator.
Recomendações relacionadas
-
As principais áreas de aplicação dos chillers refrigerados a ar
1777No mercado nacional e internacional de resfriadores de água industriais, nosso equipamento de resfriamento de água industrial mais comum pode ser dividido em dois tipos diferentes, um tipo é o resfriador de ar industrial, o outro tipo é o resfriador de água industrial...
Ver detalhes -
Quais são as diferenças entre a câmara de teste de baixa temperatura e a refrigeração?
1958A câmara de ensaio a baixa temperatura, como se pode ver pelo seu nome, significa provavelmente fazer um ensaio criogénico. É mais importante no nosso equipamento de refrigeração. A pequena série de câmara de teste de baixa temperatura diz-lhe que a refrigeração de baixa temperatu...
Ver detalhes -
Como resolver o problema do ruído do compressor no chiller industrial de baixa temperatura?
1878No chiller industrial de baixa temperatura, o funcionamento do compressor produzirá ruído, mas apenas a diferença de tamanho do som. De um modo geral, haverá alguns problemas se o ruído for muito grande. Então, como resolver o problema do ruído do compressor?
Ver detalhes -
Requisitos de temperatura para refrigeradores de implantação iónica em processos de fabrico de semicondutores
1330A implantação iónica é um passo importante nos processos de fabrico de semicondutores para dopar com precisão e controlar as propriedades eléctricas dos materiais semicondutores através da implantação de iões de alta energia nos mesmos. Este processo gera uma grande quantidade de calor...
Ver detalhes
LNEYA Refrigeradores industriais Fabricante Fornecedor