Common Faults of Salt Spray Environmental Chambers
The enterprise users are often at a loss when dealing with common faults in the salt spray environmental chamber. For this reason, Wuxi LNEYA, a manufacturer of the salt spray environmental chamber, specially reorganized the related common fault treatment technologies of the salt spray environmental chamber.
Salt spray environmental chamber for sale
The salt spray environmental chamber cannot be heated
1. Heat the solid state relay, if there is no output, replace the solid state relay.
2. If the external dial has no output, replace the external dial.
3. If there is heating tube damage, check the heating tube resistance, replace the resistance if there is something abnormal.
4. Lack of water can cause heating shielding, check whether the water level is normal or not and whether the floating ball malfunctions or not.
5. If there is a malfunction of the over-temperature maintenance system, adjust the maintenance temperature value. If the fault still cannot be solved, replace the over-temperature maintenance system.
The heating temperature is abnormal
1. If the set temperature of the temperature controller of the environmental chambers for sale is too low, set the temperature controller at the required temperature.
2. If the set temperature of the safety maintenance switch of the environmental chamber is too low, set the safety maintenance switch at the required temperature.
3. If the set temperature of the temperature controller of the full bucket is too low, set the temperature controller at the required temperature.
4. If the set temperature of the safety maintenance switch of the full bucket is too low, set the safety maintenance switch at the required temperature.
The external dial doesn’t display
1. Check whether or not there is a power supply in the salt spray environmental chamber. If not, ask a professional electrician for a check.
2. Check whether or not the power switch is damaged. If damaged, replace the power switch.
3. Check whether or not the power circuit breaker is damaged. If damaged, replace the circuit breaker.
4. Check whether or not the external dial terminal has a power supply. If there is a power supply, it indicates that the external dial is damage, replace the external dial.
5. Check whether or not the power supply wiring terminal of the salt spray environmental chamber is blackened. If it is blackened, it indicates the possibility of burning out, replace the external dial.
If the failures the enterprise users are encountered are more serious than above, it is suggested to contact the salt mist environmental chamber manufacturer to solve the failure.
Recomendações relacionadas
-
-
Descrição da pressão do refrigerador dedicado de laboratório
1571O refrigerador específico para laboratório é um dos equipamentos mais populares na indústria de equipamentos mecânicos. Por esta razão, as instruções de pressão relevantes foram especialmente compiladas para o ajudar a compreender melhor o refrigerador específico para laboratório...
Ver detalhes -
What are the temperature control performance of laboratory recirculating chillers
1719The laboratory recirculating chiller is atypical cooling circulator with heating function to ensure the repeatability oflaboratory results. There are water-cooled or air-cooled refrigerators, whichcan be placed in a laboratory fume hood or set ...
Ver detalhes -
The difference between a large refrigeration and heating circulator and a test chamber
14661. For the high temperature test under special circumstances, the high and low temperature test chamber may not meet the requirements. Generally speaking, the high temperature upper limit of the high and low temperature test box is 200 ℃, and its ...
Ver detalhes
LNEYA Refrigeradores industriais Fabricante Fornecedor