Como eliminar o alarme de alta pressão do congelador industrial de temperatura ultrabaixa?
During use, if a fault is not solved in time, it will affect the use of industrial ultra-low temperature freezer. How to solve the problem of high-pressure alarm?
If the water-cooled industrial ultra-low temperature freezer presents a high-pressure alarm, it is mainly related to the cooling water system. The water-cooled industrial ultra-low temperature freezer provides circulating cooling water through a cooling tower. If you forget to open the cooling water valve during use, If the cooling water cannot be circulated to the chiller, it will show high pressure problems. In addition, check whether the cooling water tower is defective and cannot safely and stably deliver the cooling water.
The cooling water flow is too small, or the temperature is too high and the heat dissipation is not good. This problem will also occur. If the cooling water flow is small, it is necessary to check whether the pipe size of the supporting device matches the size of the industrial ultra-low temperature freezer group, or whether the water valve can be opened to a large position; if it is an air-cooled industrial ultra-low temperature freezer, we need to look at the ambient temperature of the device If it can be exceeded, the heat must be exhausted in time to ensure that the industrial ultra-low temperature freezer unit is in an air circulating environment.
Industrial ultra-low temperature freezers are used for too long, and it is easy to accumulate dirt and other debris on the pipe wall or condenser copper pipes, which will inevitably affect the refrigeration cycle system. Therefore, we regularly provide industrial ultra-low temperature freezers according to the maintenance time recommended by the manufacturer Do maintenance cleaning work (cleaning the cooling water system, cleaning the condenser, changing the filter, etc.).
Check whether the refrigerant is excessive. The industrial ultra-low temperature freezer unit will be filled with the appropriate amount of refrigerant before leaving the factory. Therefore, if there is no leakage of the refrigerant, the user does not need to charge the refrigerant. Too much refrigerant will only affect the unit’s failure. use.
Whether the opening of the shrink valve is too small, for this, everyone can gradually adjust the opening of the shrink valve to eliminate the problem.
Recomendações relacionadas
-
Quais são os locais a escolher para os refrigerados?
1904Atualmente, com o desenvolvimento da ciência e da tecnologia, muitos clientes centram-se na utilização de temperaturas altas e baixas para melhorar a qualidade dos produtos e reduzir o seu custo. Então, quais são os lugares para prestar atenção ao escolher um modelo? O homem...
Ver detalhes -
Low temperature freezer (explosion proof) product application instructions
1253LNEYA low temperature freezer (explosion proof) is mainly used for low temperature storage in chemical explosion areas. The equipment adopts microcomputer control, multiple safety protection and more reliable operation; single compressor automatic...
Ver detalhes -
Operation method of reaction kettle for circulating heating device
1723The laboratory single-layer glass reactor equipped with the circulating heating device can place a flask containing the reaction solvent on the inner layer for stirring reaction, and the interlayer can realize constant temperature heating and cons...
Ver detalhes -
A importância do equipamento de controlo da temperatura TCU na indústria farmacêutica
15071. condições de baixa temperatura necessárias para reacções bioquímicas no interior do reator: síntese de medicamentos a baixa temperatura, etc. 2. arrefecimento rápido das matérias-primas para evitar a deterioração dos medicamentos, cristalização e extração a baixa temperatura. G...
Ver detalhes
LNEYA Refrigeradores industriais Fabricante Fornecedor