Pontos de instalação das criobombas nos frigoríficos criogénicos
Na construção do frigorífico criogénico, a bomba criogénica é indispensável. O fabricante de congeladores de baixa temperatura LNEYA diz a todos que a bomba criogénica deve ser utilizada de acordo com os requisitos aquando da sua instalação. As facilidades de instalação...
Installation facilities and requirements:
The foreline pump—the cryopump requires a pre-stage pump pre-vacuum chamber to draw the start-up pressure of the cryopump and to remove the gas generated during regeneration when the cryopump is regenerated.
Adsorption well—Install the water-cooled baffle. If the oil vapor returns to the cryopump, the low-temperature freezer tells you that if its adsorption array is contaminated, the adsorption array must be replaced.
High vacuum valve—installed between the cryopump and the vacuum chamber, which can shorten the start-up time, regenerate the cryopump without deflation in the vacuum chamber, and leak the vacuum chamber at any time.
Pre-pumping valve—At the pre-pumping port of the cryopump, a valve must be installed. The valve is open during pre-extraction and regeneration, and is closed during operation of the cryopump. Another one is used for The same mechanical pump pre-extracts the vacuum chamber.
Installation of compressor:
Make sure the compressor is placed on a stable level.
Static pressure – turn off the compressor and let it be at room temperature. The pressure should be between 1.4-1.5 Mpa. If the pressure indication is low, you need to add some He gas (99.999% purity).
The cooling water pipe connected to the compressor is connected to the water-cooled compressor.
Install the cryopump:
High vacuum flange connection
The cryostat Xiaobian tells you that the cryopump can be fixed in any place required by the vacuum chamber by means of its high vacuum flange. Please check whether the “O” ring is intact before installation, and the surface is free of indentation and debris. .
Connection of the suction port and the safety valve
The cryopump cannot be vented during operation, so it is necessary to pre-pump the cryopump through the suction port by using a rough pump such as a mechanical pump before starting, and it can be used as an outlet for the gas released during regeneration. Be sure to connect the pumping port of the cryopump to the pre-stage pump of the corresponding size.
Connect the cryopump and compressor
Remove the cryopump and compressor, as well as the protective caps on all self-sealing joints on the metal hose.
Note: Never use a cleaner to wipe the self-sealing joint in the following steps.
Before connecting, you must ensure that there are no dust, oil, particles, etc. on the joint. If necessary, use a brush to clean the thread. Use a soft cloth without cotton to wipe any dirt on the sealing surface.
Connect the metal hose, connect the hose according to the direction of the airflow, start from the compressor to the cryopump, then from the cryopump to the compressor, use two wrenches to tighten each self-sealing joint to protect the hose from twisting and leaking. , moderately tighten each joint. The low-temperature freezer, which is good, tells everyone that after the connection is completed, the indication of the pressure gauge of the compressor is checked. If it is 1.3 to 1.5 MPa, it is normal. If it is lower than this value, the purity is greater than 99.999% as described above. Add some more to your suffocation.
Special Note: The metal hose maintains a high pressure of 1.4-1.5 MPa, so it should be used with great care to prevent damage or reverse.
Recomendações relacionadas
-
Is it right to extend battery life by lowering the ambient temperature?
955The battery can work in a wide temperature range, but this does not allow the battery to be charged under high temperature conditions. The charging process is more delicate than the discharging process, and extreme care must be taken. Extreme cold...
Ver pormenores -
As principais áreas de aplicação dos chillers refrigerados a ar
1139No mercado nacional e internacional de resfriadores de água industriais, nosso equipamento de resfriamento de água industrial mais comum pode ser dividido em dois tipos diferentes, um tipo é o resfriador de ar industrial, o outro tipo é o resfriador de água industrial...
Ver pormenores -
Proteção contra explicações para o refrigerador de precisão arrefecido a ar
1045The places where need air-cooled precision chiller indicate they have high demanding for operating accuracy. Therefore, the protection settings have to work out. Then, which kind of protection for air-cooled precision chiller should be done?Discha...
Ver pormenores -
Como manter um desempenho estável do arrefecedor de óleo de laboratório biofarmacêutico?
8831. Um ambiente de trabalho adequado para evitar várias falhas do refrigerador de óleo no laboratório biofarmacêutico precisa de proporcionar um melhor ambiente operacional para o equipamento. Manter o ambiente limpo e arrumado, reduzir o impacto de vários factores...
Ver pormenores